When transmitting using my infrared device, it fails with the error message "Extraction of data from device failed”

The IR sensor of the device must be directed toward the center of the IR dongle and the device must be kept completely still during upload/transmission. Other devices in the same room with active IR interfaces might interfere with the IR connection, causing the transmission to fail. Please make sure no more IR devices are active in the same room. The IR dongle and device should be placed on a flat surface.

Please note that some devices need to be 'awake' (illuminated) in the very beginning of the upload for the cable to detect the device. The device does however not need to be lit up during the entire upload. 